What Raman & Infrared Spectroscopy got wrong, from 63 dissertations

The records describe common limitations and operational failures encountered when applying Raman and infrared spectroscopy across biological, environmental, and materials characterization tasks. Analytical workflows frequently broke down due to weak optical scattering, intense background interference from fluorescence or water, sample matrix contamination, and inferior performance compared to established analytical baselines. These records come from PhD theses at 25 institutions, 2021 to 2026. Each links to its thesis. They were extracted by language models reading the full text, so treat each as a lead to read, not a verdict.

Low analyte concentrations and weak optical scattering fall below spectroscopic detection limits

13 theses · 10 institutions

Target signals could not be detected when analyte concentrations were too dilute or when physical scattering cross-sections were inherently weak. Prolonged integration times, unenhanced measurements, and specialized detector configurations failed to resolve distinguishing spectral features for sub-nanometer defects, low-abundance minerals, and dilute biomolecules.

Tried and failed

Raman spectroscopy for defect detection applied to sub-nanometer defects in 2D materials. Outcome: no signal. Reason: optical measurement lacked sensitivity to detect sub-nanometer atomic defect formation across ion doses

Low-Dimensional Material Devices For Atomic Defect Engineering, Ionic And Molecular Transport · Penn

Tried and failed

unenhanced Raman spectroscopy with long acquisition times applied to low-concentration bacterial pathogen classification. Outcome: no signal. Reason: signal intensity at clinically relevant low concentrations was insufficient without enhancement despite prolonged integration

Raman spectroscopy for point of care urinary tract infection diagnosis · Imperial

Tried and failed

blind source separation on mixture spectra applied to low-intensity component identification. Outcome: no signal. Reason: low Raman scattering intensity prevented detection without artificial 100-fold amplification of its reference spectrum

Development of In Situ Monitoring and Data-Driven Modeling for Complex Systems: Case Study on Simulant Mixtures of Nuclear Waste · Georgia Tech

Tried and failed

low-power ambient Raman spectroscopy for phase detection applied to biphasic fluid inclusion density measurements. Outcome: no signal. Reason: low power and ambient temperature fail to detect small coexisting liquid volumes near the vapor boundary

MAGMATIC DRIVERS OF EXPLOSIVE VOLCANISM: INSIGHT FROM MAGMA MIXING, MELT INCLUSIONS AND RAMAN SPECTROSCOPY · Cornell

Tried and failed

785 nm Raman spectroscopy applied to detecting high-wavenumber O-H stretching bands. Outcome: no signal. Reason: low quantum efficiency of silicon CCD detectors near the bandgap cutoff wavelength

Advanced applications of Raman spectroscopy and machine learning techniques · Iowa State

Tried and failed

Raman spectroscopy for organic detection applied to mineral precipitate coprecipitation samples. Outcome: no signal. Reason: Organic concentrations sufficient for mineral stabilization fell below the detection limit of Raman spectroscopy.

Organic influences on hydrated magnesium carbonate mineral formation · MIT

Considered and rejected

Considered and rejected: Dismissed solid-phase FTIR and Raman spectroscopy for detecting low-level auxotrophic UAA incorporation due to insufficient sensitivity and signal attenuation

An interdisciplinary approach for developing novel methods to study silks · Oxford

Tried and failed

visible-near-infrared spectroscopy with partial least squares regression applied to low-concentration iron compounds in soil. Outcome: no signal. Reason: insufficient spectral sensitivity and absorption feature definition at low analyte concentrations (0–5000 mg/kg)

Evaluation and adaptation of portable x-ray fluorescence and data fusion techniques for non-invasive characterisation of metal-contaminated environments · Cranfield

Tried and failed

diffuse reflectance spectroscopy calibration applied to aged hydrocarbon contaminated soil. Outcome: no signal. Reason: Weathering and aging significantly reduced target analyte concentrations below spectroscopy detection sensitivity

Implementation of spectroscopy as a rapid measurement tool (RMT) to inform risk assessment at petroleum contaminated sites in the Niger Delta, Nigeria. · Cranfield

Tried and failed

surface-enhanced Raman spectroscopy on precursor controls applied to isolated nanoparticle precursor mixtures. Outcome: no signal. Reason: isolated precursors or bare nanoparticles lacked sufficient cross-section or enhancement to produce detectable spectral signals

Advances in optical trapping beyond biophysics: combining force and optical spectroscopies under diverse chemical conditions · OpenBU

Tried and failed

surface-enhanced Raman spectroscopy single-peak detection applied to sub-nanomolar analyte concentration sensing. Outcome: no signal. Reason: insufficient spectral changes at low nanomolar concentration thresholds for single-peak spectral analysis

Polypyrrole-coated gold nanoflowers as multifunctional nanoparticles for cancer theranostics · Imperial

Tried and failed

Raman spectroscopy for localized stress characterization applied to laser-micromachined silicon surfaces. Outcome: no signal. Reason: laser-induced surface modifications and shallow stress depths produced undetectable Raman frequency shifts

Ultrafast Laser Micromachining for Correction of Thin Optics for Next Generation X-ray Space Telescopes · MIT

Tried and failed

density separation and micro-FTIR spectroscopy applied to marine sediment core microplastic extraction. Outcome: no signal. Reason: Target particles were below detection limit or completely absent in specific sediment depth layers

Investigation of Microplastics from Deep-Sea Antarctic and West Coast North American Sediments · Harvard

Absence of distinct vibrational bands and spectral overlap hinder compound differentiation

11 theses · 6 institutions

Spectra frequently exhibited severe peak overlap or completely lacked diagnostic vibrational features required to distinguish chemical species, phases, or continuous ratio boundaries. Spectral deconvolution algorithms and frequency shift tracking failed when spectral responses were restricted to uniform amplitude variations or lacked sufficient degrees of freedom.

Tried and failed

Raman spectroscopy for tracking molecular doping levels applied to organic hole transport materials. Outcome: no signal. Reason: overlapping vibrational peaks between neutral molecules and polaronic charged states prevented distinction

Organic and hybrid semiconductors for photovoltaic application — control of energetics and interfacial properties · Imperial

Considered and rejected

Considered and rejected: Rejected direct femtosecond pulse excitation without spectral focusing due to inadequate spectral resolution for resolving narrow Raman fingerprint modes (<10 cm-1).

ADVANCEMENTS IN STIMULATED RAMAN SCATTERING MICROSCOPY · ResearchWorks

Tried and failed

partial least squares regression on infrared spectra applied to narrow continuous ratio class boundaries. Reason: prediction error exceeded the width of the target classification intervals

A soil organic carbon indexing and measurement system · Cranfield

Tried and failed

visible and near-infrared reflectance spectroscopy applied to subsurface sediment mobilization landforms. Outcome: no signal. Reason: Spectra lacked diagnostic absorption features required to identify expected alteration minerals like phyllosilicates, carbonates, or sulfates

Subsurface Sediment Mobilization on Mars: Insights Provided by Orbital Remote Sensing Datasets · Georgia Tech

Tried and failed

stimulated Raman scattering unmixing microscopy applied to organism digestive tract quantification. Outcome: no signal. Reason: ingested dietary material produced broad, highly variable interfering background signals

Chemical microscopy and image analysis for biomedical applications · Harvard

Considered and rejected

Considered and rejected: Rejected relying solely on FTIR spectroscopy for identifying polyethylene waste types due to lack of distinct absorption differences at 4 cm-1 resolution; adopted DSC second-heating analysis instead.

Potential transition to a circular economy for plastics in developing countries: recycling plastic waste into plastic-bonded aggregate composites for construction · Imperial

Tried and failed

ATR-FTIR spectroscopic imaging applied to identifying inorganic degradation products. Outcome: no signal. Reason: vibrational modes of these inorganic species lacked sufficient infrared activity or distinct absorption bands within measured range

Applications of advanced atr-ftir spectroscopic imaging to objects of cultural heritage · Imperial

Tried and failed

two-component spectral deconvolution modeling applied to mixed-phase infrared absorption spectra. Outcome: worse than baseline. Reason: two components lacked sufficient degrees of freedom to capture the observed multi-phase spectral variations

Infrared Spectroscopy at Three and Six Microns of Water Ices Mixtures of Astrochemical Interest · Georgia Tech

Tried and failed

infrared center-frequency shift spectroscopy applied to detecting biomolecular condensate desolvation. Outcome: no signal. Reason: vibrational lineshapes showed no appreciable frequency shifts between dilute and condensate phases despite phase separation

Gotta keep ‘em separated : probing phase separation with infrared spectroscopy · UT Austin

Tried and failed

multiphase spectral deconvolution applied to infrared vibrational absorption bands. Outcome: no signal. Reason: the spectral feature varied predominantly in amplitude rather than shape or position across phases

Infrared Spectroscopy at Three and Six Microns of Water Ices Mixtures of Astrochemical Interest · Georgia Tech

Tried and failed

polarized reflectance spectroscopy applied to interfacial water layer optical absorption. Outcome: no signal. Reason: No measurable interfacial absorption layer or predicted frequency-dependent absorption dips were observed under ambient conditions

Electron and Molecular Transport Induced by Light-Matter Interactions for Energy and Water Harvesting · Georgia Tech

Considered and rejected

Considered and rejected: Rejected infrared absorption as the detection mechanism for LPG phase change/bubbles (liquid propane did not absorb the IR wavelengths emitted, requiring a refractive scattering approach).

Development of a liquid-phase LPG MPI conversion system · Cranfield

Substrates, containers, and preparation matrices introduce optical interference and contamination

11 theses · 9 institutions

Sample vials, mounting slides, immersion media, and embedding agents introduced dominant optical bands that obscured subtle analyte signals. Adsorbed surface contaminants, displaced solvent backgrounds, and beam damage further distorted spectra and caused false compound identifications.

Tried and failed

membrane filtration surface-enhanced Raman spectroscopy applied to clinical biofluid pathogen detection. Outcome: no signal. Reason: detector saturation and spectral interference from background particulate matter in unpurified samples

Raman spectroscopy for point of care urinary tract infection diagnosis · Imperial

Tried and failed

Handheld Raman spectroscopy through glass containers applied to pharmaceutical powder in glass vials. Outcome: no signal. Reason: Container optical interference and dispersed powder prevented distinguishable spectral signal from empty vials

New Tools for Rapid Mass Spectrometric Screening · Georgia Tech

Tried and failed

Raman spectroscopy on thin exfoliated layers applied to few-layer metal oxychloride crystals. Outcome: no signal. Reason: low Raman scattering signal relative to substrate background at 1.58 eV excitation

The Influence of Defects on the Vibrational and Electronic Properties of the Quasi 1D van der Waals Magnet CrSBr · MIT

Considered and rejected

Considered and rejected: Rejected paraffin embedding because paraffin-derived Raman bands interfered with subtle plasma-induced spectral alterations

Evaluierung der Raman-Mikrospektroskopie für das Monitoring und die Bestimmung der funktionellen Gewebeeindringtiefe der Behandlung mit physikalischem Niedertemperaturplasma basierend auf primärem und patientenspezifischem Gewebematerial · Publikationssystem UB Tuebingen

Considered and rejected

Considered and rejected: Rejected measuring optical properties (photoluminescence / Raman spectroscopy) after TEM imaging due to beam damage and carbon contamination

Structure and Properties of Moiré Interfaces in Two Dimensional Materials · Harvard

Considered and rejected

Considered and rejected: Rejected using oil immersion lenses to improve Raman depth resolution because immersion oils introduce interfering spectral peaks that mask weak protein signals.

Monitoring of biotherapeutic purification by vibrational spectroscopy · Imperial

Tried and failed

multiple linear regression spectral deconvolution applied to surface-enhanced infrared absorption spectra. Reason: displaced solvent background interference and optical grating-order shifts created false negative concentration signals

Nanoplasmonic infrared biosensors enabling detection of misfolded protein biomarkers and drug screening for neurodegenerative diseases · EPFL

Considered and rejected

Considered and rejected: Rejected analyzing single-crystal Raman spectroscopy for bulk CoSb3 due to persistent surface Sb flux signals, choosing polycrystalline samples instead.

The Effects of Lattice Anharmonicity on Phase Transitions and Thermal Conductivity Through Phonon Measurements in Energy Materials · DukeSpace

Considered and rejected

Considered and rejected: Rejected silicon and quartz substrates for Raman spectroscopy because substrate background signals obscured PVAL vibrational bands, adopting gold-sputtered glass SERS substrates instead.

Deciphering the influence of local chain order on thermal transport in macromolecular matter · Imperial

Tried and failed

single-point ATR-FTIR spectroscopy applied to surface-contaminated microplastics. Outcome: no signal. Reason: adsorbed chemical contaminants obscured surface spectra causing unmatchable spectra or false compound identifications

Microplastics in Wastewater Treatment Plant Sludges: Detection Methods, Seasonal Variation, and Effects on Anaerobic Digestion · Carleton University Institutional Repository

Tried and failed

infrared microspectroscopy on glass-mounted fixed samples applied to biological tissue sections. Outcome: no signal. Reason: standard glass slides and paraffin fixation cause strong infrared absorption and spectral blockage

Exploring Potential Applications of Fourier Transform Infrared Microspectroscopy Imaging · Texas Tech

Strong water absorption and moisture emissions obscure infrared spectral bands

10 theses · 7 institutions

Liquid water and ambient water vapor produced intense absorption and emission backgrounds that masked target vibrational modes across infrared regions. These aqueous interferences attenuated optical transmission, generated baseline correction artifacts, and distorted quantitative measurements in biological tissues and aqueous suspensions.

Considered and rejected

Considered and rejected: Mid-infrared (MIR) spectroscopy was rejected for in-field proximal NUE sensing due to requirement of sample preparation, spectral distortion, and strong masking by soil water.

Proximal sensing of soil properties for nitrogen use efficiency improvement · Research Repository UCD

Tried and failed

infrared spectroscopy detection using multi-pass absorption cell applied to trace volatile organic compound detection. Outcome: no signal. Reason: the analyte absorption peak overlapped with a dominant ambient water vapour absorption band

Nanoporous Silicon in Gas Preconcentration and Sensing · Cambridge

Tried and failed

thermal infrared imaging applied to in vivo gastrointestinal inflammation detection. Outcome: no signal. Reason: optical absorption and additive emission noise from luminal gases and water vapor obscured localized thermal features

Wireless Thermal Capsule Endoscopy: A Bowel Preparation Free, Sedation Free, Outpatient Alternative to Colonoscopy · MIT

Tried and failed

reflectance photoplethysmography at shortwave infrared wavelengths applied to optical haematocrit ratio estimation. Outcome: unstable. Reason: strong water absorption and small photodiode area caused severe optical attenuation and noisy ratios

Wearable multi-wavelength photoplethysmography and algorithms for continuous non-invasive monitoring of Dengue fever · Imperial

Tried and failed

transmission mode FTIR spectroscopy applied to aqueous chromatographic resin beads. Outcome: no signal. Reason: strong water absorption and large bead dimensions prevented optical transmission

Monitoring of biotherapeutic purification by vibrational spectroscopy · Imperial

Considered and rejected

Considered and rejected: Rejected water baseline subtraction during IR/Raman preprocessing because IR water baseline is overly sensitive to probe position and agitation speed, introducing spectral artifacts.

Development of In Situ Monitoring and Data-Driven Modeling for Complex Systems: Case Study on Simulant Mixtures of Nuclear Waste · Georgia Tech

Considered and rejected

Considered and rejected: Rejected using reconstituted actin in standard H2O for 2D-IR spectroscopy due to the overwhelming background water OH vibrational band in the amide I region, substituting D2O.

In-vitro active gels: quantifying actin network interactions and morphologies · UT Austin

Considered and rejected

Considered and rejected: Decided against quantitatively interpreting the broad bridge-bound CO infrared feature due to its excessive breadth and convolution from water vapor evacuating from the catalyst layer.

The Electrochemical Nature of Non-Faradaic Catalysis at Interfaces · MIT

Considered and rejected

Considered and rejected: Rejected using mid-infrared absorption feature (~2900 cm⁻¹) for HCl CRDS due to strong water vapor interference and lower mirror reflectivity (<99.99%).

Discerning Trends in Wintertime Reactive Chlorine Chemistry and Air Quality · YorkSpace

Considered and rejected

Considered and rejected: Temperature measurements located in the gap areas between non-active and active chromium coatings were excluded/rejected from analysis due to infrared emission interference from water causing count discontinuities.

Inferencing Techniques for Enhanced Monitoring of Thermal-Fluid Systems · MIT

Fluorescence and photoluminescence backgrounds overwhelm Raman scattering signals

8 theses · 3 institutions

Excitation lasers induced intense fluorescence from organic constituents, carbon dot structures, and trace rare earth impurities that swamped weak Raman peaks. Reducing laser power left target peaks obscured by the background noise, whereas increasing power risked sample degradation without eliminating photoluminescence.

Tried and failed

visible laser Raman spectroscopy applied to ionic liquids. Reason: strong temperature-dependent fluorescence background obscured the Raman signal

Advanced applications of Raman spectroscopy and machine learning techniques · Iowa State

Tried and failed

near-infrared laser excitation Raman spectroscopy applied to narrow-bandgap semiconductor heterostructures. Outcome: no signal. Reason: laser wavelength proximity to material bandgap produced strong fluorescence background that overwhelmed Raman signal

Computational Raman imaging and thermography · MIT

Tried and failed

785 nm Raman spectroscopy applied to mineral phases in cementitious materials. Outcome: no signal. Reason: Photoluminescence from trace rare earth impurities overlapped and obscured characteristic Raman peaks

Assessing long-term carbonation resistance of blended concretes from short-term natural exposure · Imperial

Tried and failed

confocal Raman micro-spectroscopy applied to aged concrete binder phases. Outcome: no signal. Reason: impurity fluorescence overwhelmed and obscured weak characteristic Raman signals

Time-Space-Resolved Raman Analysis of Structure-Property Relationships in Heterogeneous Structural Materials · MIT

Tried and failed

1H NMR and Raman spectroscopic characterization applied to purified carbon dots. Outcome: no signal. Reason: insolubility, strong fluorescence background interference, and low mass yields

Unraveling the fluorescence puzzle: examining the role of purification and passivation in carbon dots · Imperial

Tried and failed

Raman spectroscopy applied to immature organic biological tissues. Outcome: no signal. Reason: Strong fluorescence background from high organic content overwhelmed the Raman signal

Time-Space-Resolved Raman Analysis of Structure-Property Relationships in Heterogeneous Structural Materials · MIT

Tried and failed

direct unenhanced Raman spectroscopy applied to clinical urine specimens. Outcome: no signal. Reason: broad biochemical variability and high fluorescence background masked diagnostic spectra

Raman spectroscopy for point of care urinary tract infection diagnosis · Imperial

Considered and rejected

Considered and rejected: Rejected using 1H NMR and Raman spectroscopy for MCD characterization due to poor solvation signal, background fluorescence, and excessive required sample mass.

Unraveling the fluorescence puzzle: examining the role of purification and passivation in carbon dots · Imperial

Considered and rejected

Considered and rejected: Selected a higher 4 mW laser power over a 0.8 mW setting for Raman spectroscopy because 0.8 mW yielded small, broad peaks obscured by strong fluorescence background, despite 4 mW inducing sample degradation upon repeat exposures.

The atmospheric entry of micrometeorites on Mars: Implications for their mineralogy, texture and organic constituents · Imperial

Considered and rejected

Considered and rejected: Spatial Heterodyne Raman Spectroscopy (SHRS) rejected for biological and highly fluorescent samples due to excitation wavelength limitations and single-channel constraint

Chemical Sensing as a Utility using Swept-Source Raman Spectroscopy · MIT

Unprocessed spectral models and standalone spectroscopy underperform alternative analytical techniques

8 theses · 7 institutions

Spectroscopic classification models trained on raw spectral inputs were consistently beaten by principal component reductions and manual visual inspection. Handheld or single-modality vibrational spectrometers also produced higher error rates and false positive classifications compared to chromatographic baselines or fused multi-sensor data.

Lost to a baseline

Simulated in-line Raman DBC had 16.98% %CV relative to UV chromatograms, underperforming HPLC (1.25% %CV) and well-plate Raman (1.99% %CV).

Monitoring of biotherapeutic purification by vibrational spectroscopy · Imperial

Lost to a baseline

Visual spectral comparison correctly differentiated inkjet Raman spectra via 1-3 minor peaks where automated LDA classifiers (PCA-LDA, PLS-DA, Sparse LDA) produced concordant misclassifications.

DEVELOPMENT AND VALIDATION OF A TECHNIQUE FOR ANALYSIS OF TONER PRINTED DOCUMENTS USING MAGNETIC FLUX MEASUREMENTS · DSpace at SHSU

Lost to a baseline

Raw spectral data SVM (0.722 accuracy) performed worse than 10/50 PCs SVM (0.833) on tissue block samples with portable Raman spectroscopy

Characterization of cervid skin tissues with chronic wasting disease by Raman spectroscopy and machine learning · Iowa State

Lost to a baseline

HPLC was far more specific and accurate than handheld NIR and Raman spectroscopy, which produced false positive flags on authentic/compliant high-dose formulations.

RETROSPECTIVE DESCRIPTIVE STUDY ON POST-MARKET MEDICINE QUALITY-RELATED COMPLAINTS AND THE APPLICATION OF FIELD-BASED DETECTION OF POORQUALITY MEDICINES USING NEAR-INFRARED AND RAMAN SPECTROSCOPIC SCREENING METHODS · De Montfort Open Research Archive (DORA)

Lost to a baseline

Raw spectral data ANN (0.833 accuracy) performed worse than 10 PCs ANN (0.944) on tissue block samples with portable Raman spectroscopy

Characterization of cervid skin tissues with chronic wasting disease by Raman spectroscopy and machine learning · Iowa State

Lost to a baseline

Conventional Raman achieved lower RMSEC (0.00851 vs 0.0222 for SERS) and better signal stability for phenylalanine due to its strong aromatic ring peak.

OPTIMIZATION AND ANALYSIS OF PROTEINOGENIC AMINO ACIDS USING IN-SITU SURFACE-ENHANCED RAMAN SPECTROSCOPY (SERS) AND CONVENTIONAL RAMAN SPECTROSCOPY · JScholarship

Lost to a baseline

Raman spectral imaging was marginally less precise than FT-IR in recognizing dairy residue products (accuracy of 97.51% with 14 LVs vs 98.73% with 5 LVs, respectively).

Multiscale spectral imaging for food safety with relevance to dairy processing · Research Repository UCD

Considered and rejected

Considered and rejected: Rejected single-modality spectroscopy in favor of mid-level data fusion because individual Vis-NIR (85.8%) and Raman (87.3%) models yielded inferior classification compared to fused data (92.8%).

Signature of Irish grass-fed beef - Authentication of the dietary background and geographical origin of Irish grass-fed beef using chromatographic and rapid spectroscopic techniques integrated with chemometrics · Research Repository UCD

Lost to a baseline

Quantitative dispersion spectroscopy RMSEC for Glucose (0.17 g L-1) was beaten by direct absorption spectroscopy RMSEC (0.05 g L-1)

Advances in mid-infrared dispersion spectroscopy of liquids and photothermal gas sensing · DSpace-CRIS at TU Wien

Left open by the authors

Problems the authors named and did not get to.

Left open

Perform Raman thermography on III-V materials mitigating 785 nm fluorescence using polarization selection or anti-Stokes Raman spectroscopy. Blocker: Requires physical Raman microscopy apparatus, 785 nm laser setup, and physical III-V semiconductor samples

Computational Raman imaging and thermography · MIT

Left open

Compare data-preserving versus floor-valued baselining methods on Raman urinalysis spectra to assess false-positive rates from unrelated molecular signals. Blocker: Requires private Raman spectroscopy urine datasets or access to physical urinalysis samples and spectrometer

The Applications of Raman Spectroscopy Assisted Urinalysis: Hematuria and Bladder Cancer Detection · Virginia Tech

Left open

Spectrally filter the Raman laser using Echelle gratings to rectify anomalous Raman scattering rates. Blocker: Requires optical lab hardware, a physical Raman laser system, and trapped ion apparatus.

Multi-tone Mølmer-Sørensen interaction outside of the Lamb-Dicke regime · Imperial

Left open

Improve Raman spectroscopy sensitivity and disc triggering to measure spectra of lubricant films thinner than 1.5 µm. Blocker: Requires a physical experimental elastohydrodynamic lubrication rig and Raman spectroscopy hardware setup

Effect of polymers on lubricant rheology in an elastohydrodynamic contact · Imperial

Left open

Perform high-resolution characterisation to identify unidentified Raman peaks at 1456 cm^-1 in RR2450 and 1468/1601 cm^-1 in Stellite 20. Blocker: Requires physical access to alloy samples and advanced characterisation / Raman spectroscopy laboratory equipment.

Sliding wear tests and numerical modelling of Co- and Fe- based alloys · Imperial

Left open

Integrate in situ Raman spectroscopy into continuous-flow microreactors for monitoring nanoparticle synthesis kinetics. Blocker: Requires a wet chemistry laboratory, microfluidic setup, and Raman spectrometer hardware.

Importance of Monitoring the Synthesis of Light‐Interacting Nanoparticles – A Review on In Situ, Ex Situ, and Online Time‐Resolved Studies · Cambridge

Left open

Synthesize bioorthogonal Raman probes functionalized with isotopes, azides, or triple bonds for intracellular Raman detection. Blocker: Requires a wet chemistry and cell culture lab to synthesize probes and perform intracellular Raman spectroscopy

Polypyrrole-coated gold nanoflowers as multifunctional nanoparticles for cancer theranostics · Imperial

Left open

Extend Loudon-Fleury Raman response calculations to incorporate crystalline topological defects in two- and three-dimensional quantum magnet models. Blocker: Lack of specified analytical models or tractable 2D/3D tensor network algorithms for defect-modulated Raman response

Dynamics and Transport in Strongly Correlated Quantum Magnets · Georgia Tech

Left open

Develop and evaluate multimodal optical sensing combining DRS with Raman spectroscopy, fluorescence spectroscopy, or hyperspectral DRS imaging. Blocker: Requires custom physical optical hardware, spectrometer instrumentation, and physical/biological tissue samples.

Sensing rectal cancer hypoxia for advanced prognostication and response monitoring · Imperial

Left open

Record simultaneous low- and high-dispersion Raman spectra to measure gas composition and temperature accounting for H2O and CO2 broadening. Blocker: Requires a physical experimental Raman spectroscopy lab apparatus and high-pressure gas setup

Thermometry in high pressure gases using spontaneous Raman scattering · UT Austin